Specifications: (See all)
Measuring range
-200°C to 962°C (according to ITS-90) -200°C to 850°C (according to DIN EN60751)
Permission
1 mK
Аccuracy
10 mK (optional 5 mK in the range from -50°C to 250°C)
Measurement channels
2 (optionally up to 81 with external multiplexers)
Sensors
Pt-100 sensors (optionally Pt-25) in four-wire technology
Description

The ATP 4200 is designed for accurate temperature measurement in accordance with the requirements of laboratories and quality assurance (ISO 9000, etc.). The conversion of the measured resistance value to the measured temperature value is carried out using calibration tables in the measuring device depending on the different sensor specifications. It is not necessary for the user to enter calibration data after each probe change using the keypad built into the meter. Since the calibration data of several sensors can be stored in the measuring device, in many cases it will be enough to associate the corresponding calibration data with the channel to which the new sensor was connected. If the calibration data of the newly connected proof is not yet stored in the measuring device, the user can easily compose the calibration data on the PC using the supplied software. Full compilation and maintenance of calibration tables is performed on a PC. If necessary, new calibration data can be downloaded from a PC to the ATP 4200 using this PC software.


Specifications
Main characteristics
Аccuracy
10 mK (optional 5 mK in the range from -50°C to 250°C)
Technical data
Measuring range
-200°C to 962°C (according to ITS-90) -200°C to 850°C (according to DIN EN60751)
Permission
1 mK
Measurement channels
2 (optionally up to 81 with external multiplexers)
Sensors
Pt-100 sensors (optionally Pt-25) in four-wire technology
Connectors
Lemo 1S, four-pole, on the front panel
Measuring current
approx. 0.5 mA when switching DC (approx. 1 mA for Pt-25)
Measurement time
about 1 sec. per channel, about 0.1 sec. to the channel from reduced resolution
Display
LCD, LED backlight, 2 lines ¨¤ 16 characters, height 9 mm
Long-term
≤ 5 mK/year
Temperature coefficient
≤ 1 mK/°C
Units of measurement
°C, °F, K, Ом
Interface
RS-232 interface, isolated
Memory
128 KB (optional 512 KB)
Weight
about 2.5 kg
Power supply
230 V, 50 Hz / 115 V, 60 Hz approx. 15 VA
Dimensions
260 x 80 x 240 (width x height x depth in mm)